Saturday, 9 September 2017

Pulse Pressure Variation


Pulse pressure 



is the difference between the systolic and diastolic pressure readings.


It is msured in millimeters of mercury (mmHg). 


It represents the force that the heart generates each time it contracts. 


If resting blood pressure is 120/80 mmHg, pulse pressure is 40




Pulse Pressure Variation



One of the most important concepts in resuscitation is volume responsiveness, or the ability of the cardiac output to increase in response to a fluid challenge. 

As compared to prior static measures of venous filling, such as central venous pressure (CVP) or pulmonary artery occlusion pressure (PAOP or wedge pressure), clinicians are increasingly turning to dynamic means of assessing patients’ volume responsiveness.1,2 One such measure is the pulse pressure variation (PPV).1,2



PPV is a reflection of cardiopulmonary interactions. 


As a patient breathes, both spontaneously and with mechanical ventilation, cardiac output varies. 


The more the cardiac output varies with respirations, the more likely that patient is to respond to a fluid bolus with an increase in cardiac output. 


Using this simple principle, clinicians can take advantage the common arterial line tracing to assess a patient’s volume responsiveness.



To measure the PPV in a given patient, that patient must have consistent and demonstrable cardiopulmonary interactions. This means that the patient must:



  1. Be in normal sinus rhythm
  2. Be intubated and be mechanically ventilated, making no spontaneous respiratory efforts
  3. Be ventilated with at least 8mL/kg of tidal volume
  4. Have no significant alternations to chest wall compliance, such as an open chest

If all these conditions are met, the arterial waveform can be assessed for variation in a few simple steps.
1. To make measurements easier, condense the waveform to 6.25 mm/sec from the baseline of 25 mm/s. (This is easy to do, and is located under “speed” on most monitor settings.)
Tips1
2. Identify the maximum (during inspiration) and the minimum (during exhalation) waveforms on the undulating pattern.
Tips2

3. Using the cursor, find the systolic pressure value for the largest, or maximum, inspiratory waveform, then find the diastolic pressure. Take care to use the systolic and diastolic values on the same form. The difference in these two values is your pulse pressure. The following four figures have been captured from a monitor and enlarged for detail. The first figure below shows the value for the maximum systolic of 109 mmHg.

Tips3

This figure shows the measurement of the diastolic of the maximum waveform, or 60 mmHg.

Tips4

4. Repeat step 3 for the minimum form. Here, the systolic is 91 mmHg.

Tips5

And the diastolic is 54 mmHg.
Tips6

5. These steps will give you two pulse pressures. The pulse pressure variation is calculated as:
Tips7

Therefore, in this case, the PPmax is 49, the PPmin is 37, for a mean PP of 43. This leads to a calculated PPV of 27.9%. (12/43 X 100)

Numerous studies have found that a PPV of > 12% is associated with volume responsiveness in the operating room and intensive care unit (ICU) alike. 


A meta-analysis of PPV to predict fluid responsiveness found a sensitivity and specificity of 0.89 and 0.88 respectively.3 


A meta-analysis of 22 studies focused on ICU patients had similar findings, with a pooled sensitivity of 0.88 and pooled specificity was 0.89.2 


To date, no study has specifically looked at PPV in the ED, and this is an area for future study.



The most common pitfalls in using PPV for clinical decision-making involve failing to recognize that the patient does not meet one or more of the required criteria listed above.

 If patients are breathing spontaneously, even while on assist control ventilation, they can vary their intrathoracic pressure and thereby cause exaggerated or blunted responses to the positive pressure breaths.2 


Similarly, clinicians should note that the tidal volumes used in this assessment, of > 8mL/kg,2 are larger than are otherwise used (ideally, around 6mL/kg). 


In our practice, we temporarily increase the tidal volume just for the purposes of this assessment, to ensure that the patient is seeing sufficient positive pressure ventilation to impact hemodynamics. 


However, if these criteria are met, PPV is an easy, rapid means to assess a critically ill patient’s likelihood of volume responsiveness.

A 74-year-old lady with a history of ischaemic heart disease and severe congestive cardiac failure is admitted to the ICU with hypotension and presumed sepsis

A17

A 74-year-old lady with a history of ischaemic heart disease
and severe congestive cardiac failure is admitted to the
ICU with hypotension and presumed sepsis. She is sedated
and ventilated in pressure support mode. On examination
she is confused, BP is 85/35mmHg, HR is 115bpm (sinus
tachycardia), SpO2 is 95% on 60% oxygen. Arterial blood
gas analysis shows a lactate of 4.3mmol/L (39mg/dL).
Which is the BEST guide to the need for further intravenous
fluid replacement?


a. Response of oesophageal Doppler to passive leg raising.
b. Insertion of a pulmonary artery catheter and pulmonary artery
occlusion pressure measurement.
c. Titrate fluid resuscitation against repeated blood lactate
measurements.
d. Assess pulse pressure variation.
e. Urine output measurement.


A


This patient may have hypotension secondary to a variety of causes,
including sepsis and decompensated cardiac failure. 


Pulmonary artery 
occlusion pressure has been shown to be a poor predictor of whether a
fluid bolus will cause an increase in cardiac output (i.e. whether the patient
is volume-responsive). 


Blood lactate is a sensitive marker of shock and
falling levels correlate with improved survival.



 However, this will not
distinguish between cardiogenic shock (where inotropes may be required)
and septic shock (requiring fluids and/or vasopressors). 



Pulse pressure
variation of >13% has been shown to accurately predict response to fluid,
but is only reliable in mechanically ventilated patients without spontaneous
respiratory effort. 



Urine output measurement will again not distinguish
between different causes of shock. 



Passive leg raising autotransfuses
about 300ml of blood into the central circulation. 



If stroke volume
increases significantly (>10% as measured by oesophageal Doppler), this
indicates preload-responsiveness. 



An advantage of this technique is that it
is reversible if no improvement is seen, avoiding worsening pulmonary
oedema in the case of cardiogenic shock.

Wednesday, 6 September 2017

A15 All the following increase the likelihood of a patient
acquiring an antimicrobial-resistant infection EXCEPT:
a. Use of cefotaxime.
b. High nursing workload.
c. Prolonged mechanical ventilation.
d. Brief hospital admission.
e. Understaffing in the ICU.


D
Multi-drug resistant infections are increasing in incidence in the intensivecare unit. Use of broad spectrum antibiotics such as third generationcephalosporins, carbapenems and quinolones is an independent riskfactor for ventilator-acquired pneumonia with a drug-resistant organism, asis prolonged mechanical ventilation.

Methycillin-resistant Staphylococcus aureus rates have been shown to increase with bothovercrowding andunderstaffing of the ICU.


Prolonged hospital admission, the presence ofindwelling devices and poor hand hygiene also contribute to thedevelopment of antibiotic-resistant infections.
A14 A 48-year-old woman is rescued from a house fire during
which she was trapped in a smoke-filled bedroom for 30
minutes. On arrival in the Emergency Room, she has
marked facial burns and a hoarse voice but no stridor. She
is expectorating carbonaceous sputum, appears confused
and has a cherry-red visage. Which statement is FALSE?
a. Early intubation is advisable.
b. A significant thermal injury to the trachea is likely.
c. Lavage with sodium bicarbonate 1.4% has a role in the
management of this patient.
d. Lung function is likely to worsen over the next 12 hours.
e. A cherry red visage has several causes other than carbon monoxide
poisoning.



A14 B
Inhalational injury has three mechanisms: thermal injury to the upper
airway, smoke inhalation causing chemical pneumonitis to the lower
airways, and systemic absorption of toxins (principally carbon monoxide
and cyanide).

Hoarseness and facial burns are signs of upper airway
thermal injury to the pharynx and glottic area.

Pharyngeal oedema is likely
to increase rapidly especially once fluid resuscitation is commenced, and
early intubation is advised.

Most of the heat from hot gas inhalation is
dissipated in the upper airways, so thermal injury below the glottis is
unusual. 

Sodium bicarbonate lavage of the bronchial tree may be
performed following intubation to neutralise acidic deposits and remove
soot contamination, although evidence for the effectiveness of this therapy
is lacking.

Lung function usually worsens over a period of hours following
inhalational injury.

A cherry red visage is a non-specific sign of carbon
monoxide poisoning, and there are many other causes of facial flushing
including alcohol, emotion and heat, all of which are associated with burns
injuries.
A13 Which of the following statements regarding the use of
antifibrinolytic agents is FALSE?
a. Tranexamic acid is a competitive inhibitor of plasminogen and
plasmin.
b. Aprotinin significantly reduces blood loss and transfusion
requirements in cardiac surgery.
c. Use of antifibrinolytics in trauma is supported by several high quality
randomised controlled trials.
d. Arterial and venous thrombosis are uncommon complications of
tranexamic acid use.
e. The risk of anaphylaxis with aprotinin is 0.5%.


A13 C

  • Tranexamic acid (TXA) is a competitive inhibitor of plasmin and plasminogen, while aprotinin forms irreversible complexes with a variety of proteases including plasmin. 
  • While TXA is synthetic, aprotinin is isolated from bovine lungs and has a high incidence of anaphylaxis (0.5%). 
  • A strong evidence base supports the use of antifibrinolytic agents in cardiac surgery, where they have been shown to reduce the requirement for blood transfusion by about 30%. A Cochrane review 1 failed to demonstrate any increased incidence of thrombosis with antifibrinolytics, although a recent study 2 suggested that aprotinin is associated with an increased incidence of myocardial infarction, stroke and renal failure, leading to its withdrawal in the USA. 
  • The rationale for use in trauma is extrapolated from evidence in elective cardiac surgery; the ongoing CRASH (Clinical Randomisation of an Antifibrinolytic in Significant Haemorrhage) II trial should help determine whether they are effective in this context.

The following are prerequisites for the use of recombinant factor VIIa in bleeding trauma patients EXCEPT:

The following are prerequisites for the use of recombinant factor VIIa in bleeding trauma patients EXCEPT:
a. Platelet count >50x109/L.
b. Temperature >36°C.
c. Fibrinogen >0.5g/L.
d. pH >7.20.
e. Ionised Ca2+ >0.8mmol/L (3.2mg/dL).


B

  • ·        Recombinant factor VIIa (rFVIIa) may be helpful to induce coagulation in areas of diffuse small vessel coagulopathic bleeding.
·        It follows that all surgical avenues to control bleeding must first be explored.
·        Adequate platelet numbers are required to generate the ‘thrombin burst’ which the rFVIIa provokes, and adequate fibrinogen must be present to translate this thrombin generation into clot formation.
·        Correction of severe acidosis, hypothermia and hypocalcaemia are also warranted to maximise the chances of success for this expensive and unlicensed therapy.

·        However, a temperature of 36°C is difficult to achieve in such patients; 32°C is considered an appropriate minimum threshold.

Thursday, 19 January 2017

FEATURES INTENDED TO PREVENT HUMAN ERROR

FEATURES INTENDED TO PREVENT HUMAN ERROR
·        The medical gas pin index and diameter index safety systems ensure that medical gas connections are made correctly.
·        Therefore, one cannot hang a nitrous oxide cylinder in an oxygen hanger yoke or connect a nitrous oxide hose to the oxygen pipeline inlet of the machine.
·        The “fail-safe” valve is a pressure-sensitive device that interrupts flow of all hypoxic gases on the machine to their flow control valves if the supply pressure of oxygen in the intermediate-pressure system (i.e., components downstream of the first stage oxygen regulator that reduces the high pressure in the tank to 45 pounds per square inch gauge [psig] and upstream of the oxygen flow control needle valve) falls below a threshold (˜20 psig in GE machines; 12 psig in Dräger machines).
·        When the pressure of oxygen in the intermediate-pressure system falls below 30 psig, an oxygen supply pressure failure alarm is annunciated.
·        The oxygen flow control knob is “touch coded”; that is, it is fluted and larger in diameter than the other gas flow control knobs and is normally located on the right of all other gas flow control valves.
·        In the Dräger Fabius GS workstation, the gas flow controls are arranged vertically, with the oxygen flow control knob as the lowest.
·        Key-fill systems for anesthesia vaporizers are safety features that also decrease the likelihood of atmospheric contamination during vaporizer filling.
·        The most important of such systems is the Safe-T-Fill system used on desflurane bottles and desflurane vaporizers because filling a vaporizer specific for another agent with desflurane could result in a lethal overdose of desflurane.
·        A pressure relief valve built into the machine common gas outlet, breathing system, or ventilator provides some protection against positive pressure barotrauma.

FEATURES TO CORRECT FOR USE ERROR
·        Gas flow proportioning systems ensure a minimum oxygen concentration of 25% when nitrous oxide and oxygen are being used.
·        Therefore, if the anesthesiologist were to accidentally attempt to increase the flow of nitrous oxide, either the oxygen flow would be increased automatically or the flow of nitrous oxide would be limited according to the flow of oxygen that was set.
·        On older machines, during use of the anesthesia ventilator, changes in fresh gas flow, inspiratory-expiratory (I:E) ratio, or respiratory rate cause changes in delivered tidal volume that might result in overventilation, underventilation, or even barotrauma.
·        On modern systems (e.g., GE workstations that incorporate a ventilator with the Smart Vent feature, Dräger Fabius GS, and Apollo workstations), once the ventilation parameters have been set, they are maintained because the ventilator or circuit automatically compensates for changes in gas flow settings.
·        In the GE workstations with Smart Vent, tidal volume is monitored continually by a computer.
·        If measured tidal volume changes from that set to be delivered, the computer adjusts the volume delivered by the ventilator bellows.
·        Dräger and Datascope Anestar workstations use fresh gas decoupling to maintain a constant tidal volume.
·        In this system, during the inspiratory phase, a decoupling valve closes so that fresh gas entering the breathing system is directed into the reservoir bag and only gas from the ventilator is delivered to the patient.
·        During exhalation, the ventilator chamber refills from the fresh gas flow and the fresh gas that was collected in the reservoir bag during the previous inspiration.
·        A vaporizer interlock system prevents the unintentional simultaneous use of more than one vaporizer.

MONITORING SYSTEMS
·        A monitor of oxygen in the gas delivered to the patient is mandatorily enabled, and the low oxygen concentration alarm is
activated whenever the anesthesia workstation is capable of delivering an anesthetic gas mixture from the common gas
outlet.
Other monitors in the anesthesia workstation include pressure, volume, flow, and gas composition. Some also incorporate
airway gas flow monitoring. The breathing system lowpressure monitor alarm is automatically enabled when the ventilator
is turned on.




ALARM SYSTEMS
Contemporary workstations incorporate an integrated prioritized alarm system with visible and audible alerts when set
parameter limits are exceeded.
An important safety feature of all modern machines/workstations is the preuse checkout. In 1993, the U.S. Food and Drug
Administration (FDA) published anesthesia apparatus checkout recommendations. The machine should be checked by
an educated user. Item no. 1 on the FDA checklist is that an alternative means to ventilate the patient's lungs should be
present and functioning. Therefore, if a problem arises with the machine, the patient's lungs can be ventilated using a selfinflating
resuscitation bag (e.g., Ambu bag). If a machine problem arises and the cause/remedy is not immediately obvious, one should instinctively reach for the resuscitation bag and call for help. A recent analysis of the American
Society of Anesthesiologists Closed Claims Project data found that 35% of adverse outcomes were likely preventable if a
proper preuse checkout had been performed.
Recognizing that not all of the FDA 1993 checkout recommendations can be applied to many of the contemporary
workstations, in 2008, the American Society of Anesthesiologists published guidelines applicable to all anesthesia
delivery systems so that individual departments can develop their own workstation-specific preuse checkout that can be
performed consistently and expeditiously. The 2008 guidelines are intended to provide a template for developing checkout
procedures that are appropriate for each individual anesthesia machine design and practice setting. They discuss which
systems and components should be checked, the checkout interval (e.g., before first case vs. before every case), and who
may be responsible for performing each checkout procedure—the anesthesiologist or technician (Table 59.1). Examples

of user-developed workstation-specific checkouts are available on the American Society of Anesthesiologists'